7064 High Pressure Calibrator
7064 High Pressure Calibrator

The 7064 Features a 4.5 digit display with 5 selectable units.
External line pressure (pneumatic) is connected to the console rear panel.
Output pressure is controlled by a non - relieving type regulator, with Bleed and Block needle valves used to vent and isolate the output port. All connections are Minimess high-pressure type.

Fully integrated with the 7051 Control Centre and EasyCal the user can define pressure calibration procedures.

Specification
Range
35
70
100
200
Resolution (bar)
0.01
0.01
0.01
0.01
Regulator
Yes
Yes
Yes
Yes
Acc % +/- 1 digit
0.04
0.04
0.04
0.1
Temp Stability Less than 70 ppm per degC
Units Bar, PSI, kPa, Mpa, inWg, cmWg, inHg, mmHg, Kg/cm2, atm**
Maximum Pressure 1.5 x range
Sensor Piezo-resistive – stainless steel diaphragm
Over press warning 1.2 x range full scale – audio and visual (on LCD) warning
Fittings Minimess
  ** Specify on ordering: 5 pressure units max
General Information and Options
Panel size Width 200 x Height 201mm
Order code 7064 / Pressure range option code
Pressure Options 7105 = 35bar range. 7106 = 70bar range.
7107 = 100bar range. 7108 = 200bar range.
7080/81 Bench and Consoles
 
7082 Mains Power Control
7083 Dual 24V DC Supply
7088 Variable Power Unit
7086 Adjust. DC Power Supply
7087 Quad DC Power Supplies
 
7051 MF Calibrator/Control
EasyCal Software
 
7062 Pressure Calibrator
7064 High Pressure Calibrator
7065 Cal. Press. Indicator
7038 Multif. Press. Indicator
7066 Differential Press. Cal
7084 Pneu. Line Press. Control
7032 Auto Pressure Calibrator
 
7085 Temp. Distribution Panel
7079 Loop Anc. Calibrator
7068 RTD Calibrator
7067 Loop Calibrator
7069 V-I Loop Calibrator
7061 Multifunc. Proc. Calibrator
 
7070/72 Dry Block Calibrators
Calibration Pumps & Test Kits
Fittings and Extras
Click here for the CalBench Quote Form
Introduction
QuickView Datasheet (pdf)
Full Datasheet (pdf)
Cal. Bench Module Viewer
Bench Ordering Information