The 7032 is a low cost
programmable pressure calibrator module that features a
4.5 digit display with 5 selectable pressure units. It is
designed to speed up routine calibration of pressure gauges,
indicators and pressure transmitters. The user presets the
required calibration points, usually 0%, 25%, 50%, 75%,
100% of full scale – these are generally considered
most useful. However for example, 0%, 10%, 50%, 70%, 95%
could be used.
The
output line can be vented either by pressing the vent button
or selecting zero on the regulator switch to enable zero
checking. Output pressure is displayed on the 4.5 digit
LCD display.
Manual
operation
The 7032 can be controlled manually from a front panel,
which allows immediate selection of any of the pre-set calibration
pressures.
PC
(7051) controlled operation – EasyCal
The 7032 is connected to the 7051 Console Controller and
automatic calibration is possible. Using EasyCal, Time Electronics
universal calibration software, the calibration procedure
can be fully automated. Procedures can be written for pressure
gauges, indicators, and pressure transmitters. The results
of the calibration are automatically stored on the 7051,
and can be printed on demand. Closed loop calibration of
pressure transducers can be fully automated by implementing
transmitter read-back using the 7051’s internal DMM.
| Specification |
| Range
|
Vacuum |
0.2 |
2 |
10 |
| Resolution
(bar) |
.001 |
0.01mb |
.0001 |
0.001 |
| Regulator
|
Yes |
Yes |
Yes |
Yes |
| Acc
% +/- 1 digit |
0.04 |
0.1 |
0.04 |
0.04
|
| Temp
Stability |
Less
than 70 ppm per degC |
| Units |
Bar,
PSI, kPa, Mpa, inWg, cmWg, inHg, mmHg, Kg/cm2, atm** |
| Maximum
Pressure |
2
x range |
| Sensor |
Piezo-resistive
– stainless steel diaphragm |
| Over
press warning |
1.2
x range full scale – audio and visual (on LCD)
warning |
| Fittings |
Minimess
(>20 bar) & Quick Release (<20 bar) |
| |
**
Specify on ordering: 5 pressure units max |
| General
Information and Options |
| Power
Supply |
Internal
to Calibration Bench Console |
| Connections |
Quick
release couplings |
| Panel
size |
Width
340 x Height 201mm |
| Order
code |
7062
/ Pressure range option code |
| Pressure
Options |
7100
= 200mbar range. 7101 = 2bar range.
7103 = 10bar range. 7111 = Vacuum regulator. |