The
7005 is a micro-processor based handheld process control instrument
for voltage and current loop calibration and simulation. It
can operate in three modes:
a) Loop current/voltage Source (transmitter simulation and loop
supply simulation)
b) Sink of loop current (transmitter simulation)
c) Measurement of loop current/voltage (loop indicator simulation).
Manual
step of the output is available at five calibration points,
0%, 25%, 50%, 75% and 100% of span. Automatic stepping of the
output is also available both up and down with programmable
dwell times.
Continuous
up/down ramping is also available with user programmable ramp
rates and dwell time (top and bottom).
In source
mode the range can be user programmed to any value between 0mA
and 50mA, or 0V and 21V. For example a low point of 10mA and
a high point of 50mA could be set giving a span of 40mA
Measure
mode provides both voltage and current measuring capability
with 5 digit resolution. Ranges are 0 to ±5V and ±5
to ±25V, 0 to ±25mA and ±25 to ±125mA.
Alternatively the signal can be measured as a % of span for
the following ranges, 4 to 20mA, 0 to 20mA, square root 4 to
20mA, square root 0 to 20mA. For all measurements a Min/Max
recording function is available on demand.
An internal
NiCad battery can power the unit for typically 9 hours and an
external mains charger is supplied as standard. An automatic
power-down feature is incorporated to conserve battery life.
This process
calibrator is one of Time Electronics' comprehensive process
control equipment line. With source and measure capabilities
it delivers accuracy and reliability in one instrument.
|