The
1006 is an accurate low cost millivolt source suitable for voltage
injection applications.
Three
output ranges are provided to give adjustable output values
from 1µV to 1V with a basic 0.02% accuracy.
For signal
injection, the operator needs to switch on, check the battery
condition, select the range, and set the required voltage using
the thumbwheel switches.
The 1006
uses a precision reference diode and low temperature co-efficient
resistors to give a highly stable output.
The 1006
has up to 20 mA drive current and is short circuit and overload
protected. An off/normal/reverse output polarity switch is provided.
As an
accurate millivolt source, this well-proven portable test instrument
can be used for many applications including thermocouple simulation,
chart recorder calibration, A/D converter and DMM calibration,
and as a stable voltage for backing off DC offsets. |